Call Us: +91-7972969901
Call Us: +91-7972969901
| Main unit | 1 pc |
| USB cable and Software | 1 pc |
| 9v battery | 1 pc |
| Main unit | 1 pc |
| RS232 Cable and software | 1 pc |
| Output Connector | 1 pc |
| Battery (AAA) | 6 pcs |
| Main unit | 1 pc |
| Battery(AAA) | 2 pcs |
![]() |
![]() |
|
Measuring principle
|
Stylus method
|
|---|---|
|
Probe
|
BFW skidless system
|
|
Measuring range mm
|
+/- 250 µm (up to +/- 750 µm with 3x probe arm length)
|
|
Profile resolution
|
Measuring range +/- 250 µm: 8 nm
Measuring range +/- 25 µm: 0.8 nm |
|
A filter according to the ISO/JIS
|
Gaussian filter as per ISO 11562
A filter as per ISO 13565 |
|
Number n of sampling length according to ISO/JIS
|
1-5
|
|
Contacting speeds
|
0,2 mm/s; 1,0 mm/s
|
|
Measuring force (N)
|
0.75 mN
|
|
Weight drive unit
|
approx. 0.9 kg
|
|
Weight measuring instrument
|
approx. 1.0 kg
|
|
Surface parameters
|
Over 50 surface parameters for R, P and W profiles according to current ISO/JIS or MOTIF standards (ISO 12085)
|
Machine building
Automotive industry
Steel industry
Medicine
Measuring stand
Other accessories
![]() |
![]() |
|
|
Measuring principle
|
Stylus method
|
|---|---|
|
Probe
|
Inductive skidded probe
|
|
Measuring units
|
μm/μinch selectable
|
|
Measuring range mm
|
0,350
|
|
Profile resolution
|
8 nm
|
|
Vertical Scale
|
automatic/selectable
|
|
Horizontal scale
|
Depends on cutoff
|
|
Log contents
|
R profile, MRK,
P profile (MOTIF), results |
|
Filter according to ISO/JIS
|
Gauß-Filter, Ls-Filter
|
|
Short stroke under ISO/JIS
|
Selectable
|
|
Traversing length according ISO 12085 (MOTIF)
|
1 mm, 2 mm, 4 mm, 8 mm, 12 mm,16 mm
|
|
Evaluation lenth In according to ISO/JIS
|
1.25 mm, 4 mm, 12.5 mm
|
|
Number n of sampling length according to ISO/JIS
|
Selectable: 1-5
|
|
Contacting speeds
|
0.5 mm/s
|
|
Stylus
|
2 µm
|
|
Measuring force (N)
|
0,00075
|
|
LC display
|
High resolution color display, 3.5″, 320 x 240 pixel
|
|
Printing
|
Automatic/manual
record with time |
|
Printer
|
Thermal printer, 384 points/horizontal line, 20 characters/line
|
|
Printing speed
|
Approx. 6 lines/second corresponds to approx. 25 mm/s (1 in/s)
|
|
Thermopaper
|
Ø 40.0 mm -1.0 mm, width 57.5 mm-0.5 mm, coated
|
|
Calibration function
|
Dynamic
|
|
Storage capacity
|
Internal memory for max. 40000 results, max. 30 profiles
|
|
Password protection
|
Yes
|
|
Lock for device settings
|
Yes
|
|
Power management
|
Yes
|
|
Interfaces
|
Drive unit, power sorce, USB, MarConnect
|
|
Relative humidity
|
30 % to 85 %
|
|
System of protection
|
M 300 = IP 42, RD 18 = IP 40
|
|
Operating temp. range
|
+5 °C to +40 °C
|
|
Temperature range for storage
|
-15 °C to +55 °C
|
|
Power supplied
|
NiMH battery, capacity: approx. 500 measurements (depends on number and length of record printouts)
|
|
Rechargeable batteries
|
NiMH battery, capacity: approx. 500 measurements
|
|
Wide range power supply
|
Plug-in power pack with three power adapters, for input voltages from 90 V to 264 V
|
|
Dimensions (L x W x H) for drive unit in mm.
|
139 x 26 mm
|
|
Weight drive unit
|
Approx. 300 g
|
|
Dimensions (L x W x H) for measuring instrument in mm.
|
190 x 140 x 75 mm
|
|
Weight measuring instrument
|
Approx. 1 kg
|